Why grades matter
Glove boxes and listings use “exam”, “industrial”, “food-service”, “general purpose” and “medical grade” loosely. Only some of these words have a regulatory meaning, and the products behind them are not interchangeable. Under FDA’s rules, the words used to sell a glove help define its intended use, so a glove sold for patient care has to be cleared for it.
Exam (patient examination) gloves
An exam glove is a medical device: a disposable glove “worn on the examiner’s hand or finger to prevent contamination between patient and examiner” [1]. FDA classifies it as Class I and requires a 510(k) [1][2], and powdered patient examination gloves have been banned since January 2017 [3]. What proves exam grade is a K number whose FDA record matches the product, with an exam glove product code such as LZA for nitrile, LYY for latex or LYZ for vinyl [4]. On a GTDI quote, the K number is on the quote line.
Surgical gloves
Surgeons’ gloves are a separate device type, worn by operating room personnel to protect a surgical wound, with their own product code (KGO) [4][5]. FDA’s minimum pinhole quality for them is stricter: AQL 1.5, compared with 2.5 for exam gloves [6]. A sterile exam glove is still an exam glove, not a surgical glove.
Industrial and general-purpose gloves
“Industrial grade” has no FDA meaning, and “general purpose” has no defined meaning at all. Such gloves are not cleared medical devices unless a 510(k) exists for them. Be careful with industrial gloves whose product pages quote values from the medical nitrile specification, ASTM D6319. Citing a medical standard does not make a glove an FDA cleared exam glove [7]. For patient care, choose an exam glove.
Food-service (food-contact) gloves
Food-contact compliance is about the glove’s materials under FDA’s food-additive regulations. Nitrile and latex glove makers commonly cite 21 CFR 177.2600 [8]. FDA places responsibility for that compliance on the manufacturer [9]. It is not a medical clearance, and it says nothing about pinhole quality or suitability for patient care.
“Medical grade”
“Medical grade” is not a regulatory term. Look for the exam glove 510(k) instead.
“Chemo-tested”
A chemotherapy-tested glove has been tested for permeation by specific drugs under ASTM D6978, with a breakthrough time for each drug, and the claim is cleared through a 510(k) with product code OPJ [4][10]. It applies only to the drugs on the manufacturer’s list, and some lists carry “do not use” cautions for particular drugs [11]. Nothing rates a glove for chemotherapy. A glove is tested, and the list of drugs is the result.
Workplace rules
Where hand contact with blood or other potentially infectious materials is reasonably anticipated, OSHA’s bloodborne pathogens standard requires gloves and prohibits washing disposable gloves for reuse [12].
Questions that tell the grades apart
- What is the K number, and does FDA’s record show an exam glove product code?
- Does the box say “patient examination glove”?
- For a food-service glove, which regulation does the manufacturer state compliance with?
- For a chemotherapy claim, may I see the tested drug list and breakthrough times?
How GTDI lists gloves
Every glove is listed under its documented use. Exam gloves are listed as FDA 510(k) cleared patient examination gloves, with the K number on your quote. Industrial and food-service gloves have their own group, marked as non-medical, with the manufacturer’s food-contact statement or test reports available with the quote. A glove that holds an exam glove 510(k) is listed with the exam gloves.
Sources
- 21 CFR 880.6250, Non-powdered patient examination glove (eCFR, current as of Oct 1, 2026).
- FDA, “Medical Gloves” (content current as of Jul 2, 2024).
- Federal Register 81 FR 91722, Banned Devices; Powdered Surgeon’s Gloves, Powdered Patient Examination Gloves (Dec 19, 2016; effective Jan 18, 2017).
- FDA Product Classification database, records LZA, LYY, LYZ, OPJ and KGO (live database; accessed Oct 4, 2026).
- 21 CFR 878.4460, Non-powdered surgeon’s glove (eCFR, current as of Oct 1, 2026).
- 21 CFR 800.20 (eCFR, current as of Oct 1, 2026).
- An industrial nitrile glove product page citing ASTM D6319 values without a 510(k) (undated page; accessed Oct 4, 2026).
- 21 CFR 177.2600, Rubber articles intended for repeated use (eCFR, current as of Oct 1, 2026); manufacturers’ nitrile glove data sheets on file with GTDI.
- FDA, “Determining the Regulatory Status of Components of a Food Contact Material” (updated Feb 21, 2018).
- ASTM D6978-05(2023), Resistance of Medical Gloves to Permeation by Chemotherapy Drugs (store page updated Sep 29, 2023; scope summary).
- FDA-hosted 510(k) summaries for chemotherapy-tested exam gloves (cleared Oct 26, 2020 and Jul 20, 2026) carrying cautions against two drugs.
- 29 CFR 1910.1030, Bloodborne pathogens (eCFR, current as of Oct 1, 2026).
This guide is general information for supply planning. Your clinical, safety and compliance teams set your facility's policies. Reviewed October 5, 2026.